Positron Emission Tomography (PET) is a medical imaging modality that utilizes positron-emitting isotopes, such as Ga-68 and F-18, for many diagnostic purposes. The positron annihilates with an electron from the surrounding area, creating two photons of 511 keV energy and opposite momenta, entangled in their orthogonal polarizations. When each photon undergoes a Compton scattering process, the difference of their azimuthal scattering angles reflects the initial orthogonality of their polarizations, peaking at $\pm$90$^{\circ}$. This type of correlation is not yet utilized in conventional PET scanners, but could potentially offer an energy-independent method for background reduction. Measurements of these kinds of correlations can be achieved using Compton polarimeters, built from a single layer of segmented scintillating crystals such as Gadolinium Aluminium Gallium Garnet doped with Cerium (GAGG:Ce), read out by silicon photomultipliers (SiPMs). In this paper, we study the signal-to-random background ratios in measurements of these correlated annihilation photons from coincidence time spectra across clinically relevant source activities, from $\sim$200 MBq to $\sim$378 MBq. These are then compared to the standard single-pixel (photoelectric) measurements. We find that the signal-to-random background ratios (SBRs) obtained from the polarization-correlated events for Compton scattering angles $θ_{1,2}\in[72^{\circ}, 90^{\circ}]$ and azimuthal angle difference $ÎÏ=90^{\circ}\pm20^{\circ}$ are consistently higher than those from single-pixel events, with the ratio of their SBR values of 1.23. The SBR of the selected events also increases with the polarimetric modulation factor $μ$, gaining $\sim$50\% in value during the experiment.
academic- 论文ID: 2510.11504
- 标题: Towards polarization-enhanced PET: Study of random background in polarization-correlated Compton events
- 作者: Ana Marija Kožuljević, Tomislav Bokulić, Darko Grošev, Siddharth Parashari, Luka Pavelić, Marinko Rade, Marijan Žuvić, Mihael Makek
- 分类: physics.ins-det physics.med-ph
- 发表时间: October 17, 2025
- 论文链接: https://arxiv.org/abs/2510.11504
正电子发射断层扫描(PET)是一种利用正电子发射同位素(如Ga-68和F-18)进行诊断的医学成像技术。正电子与周围组织中的电子湮灭,产生两个511 keV能量、动量相反且偏振正交纠缠的光子。当每个光子发生康普顿散射时,其方位散射角差反映了初始偏振的正交性,在±90°处达到峰值。这种相关性尚未在传统PET扫描仪中得到利用,但可能为背景抑制提供独立于能量的方法。通过康普顿偏振计可以测量这些相关性,特别适合使用与传统PET架构兼容的单层偏振计。本研究在200-378 MBq的临床相关源活度范围内,比较了偏振相关康普顿事件与传统单像素击中事件的信号与随机背景比(SBR),发现前者的SBR始终更高。
传统PET成像依赖于湮灭光子的511 keV能量和相反动量进行成像,但由于有限的时间窗口,存在检测到两个不相关光子的随机符合事件问题,这些事件不对应真实的湮灭位点,从而降低图像质量。
- 量子纠缠利用:湮灭光子的偏振纠缠特性提供了额外的信息维度,可用于背景抑制
- 临床需求:提高PET成像的信噪比对临床诊断具有重要意义
- 技术创新:开发偏振增强PET技术代表了医学成像的新方向
传统PET仅利用能量和动量信息,未充分利用量子纠缠特性,导致随机符合背景限制了成像质量。
探索利用湮灭光子偏振相关性进行背景抑制的可行性,为开发下一代PET系统提供理论和实验基础。
- 首次详细对比研究:系统比较了单像素击中事件与偏振相关康普顿事件的随机背景特性
- 临床相关活度验证:在200-380 MBq临床相关活度范围内验证了偏振增强PET的有效性
- SBR与偏振调制因子关系:首次发现并量化了信号与随机背景比与偏振调制因子μ的相关性
- 背景抑制优势验证:证明偏振相关事件的SBR比传统单像素事件高出23%
探测器模块:
- 两个16×16的GAGG:Ce晶体矩阵
- 晶体尺寸:3×3×20 mm³
- SiPM读出(Hamamatsu S13361-0808A)
- 能量分辨率:8.1±1.1% 和 9.3±2.2% (511 keV)
几何配置:
- 探测器间距:430 mm
- 旋转机架:12个位置,间隔15°
- Ga-68源置于中心
读出系统:TOFPET2 ASIC系统
- 符合触发:每个模块至少一个像素触发
- 时间窗口:100 ns事件分组
- 符合时间分辨率:±2.5 ns (±3σ)
单像素事件:
- 每个探测器模块仅一个像素触发
- 全能量沉积:400-600 keV
偏振相关康普顿事件:
- 每个模块两个像素触发
- 总能量:400-600 keV
- 最小单像素能量:120 keV
康普顿散射角:
θ=arccos(Epx1+Epx2mec2−Epx2mec2+1)
方位散射角:
ϕ=arctan(ΔxΔy)
偏振调制因子:
μ=G(θ1)G(θ2)
其中 G(θi)=(2−cosθi)2sin2θi
根据Pryce-Ward理论,偏振相关康普顿散射的微分截面为:
dΩ1dΩ2d2σ=16r04[F(θ1)F(θ2)−G(θ1)G(θ2)cos(2Δϕ)]
方位角差分布经接受度修正后拟合:
Ncor(ϕ1−ϕ2)=M[1−μcos[2(ϕ1−ϕ2)]]
- 同位素:Ga-68
- 初始活度:~378 MBq
- 测量范围:200-380 MBq
- 衰减过程:实验持续约1小时,利用自然衰减获得不同活度
- 位置数:12个,每个位置15°间隔
- 采集时间:240-447秒(随活度调整)
- 能量窗口:400-600 keV (全能峰)
- 最小像素能量:120 keV
- 角度选择:
- 主要样本:72° < θ < 90°
- 对比样本:60° < θ < 80°
- 拓扑选择:
- 方位角选择:
- 高相关区:70° < Δφ < 110°
- 低相关区:-20° < Δφ < 20°
SBR优势:
- 偏振相关事件的SBR始终高于单像素事件
- 最佳选择(72° < θ < 90°, 70° < Δφ < 110°)的SBR比值保持在1.1以上
- 在最高活度下,SBR比值达到1.23
活度依赖性:
- 随源活度降低,SBR显著提升
- 偏振调制因子μ在活度从
387 MBq降至200 MBq时增加~28% - 排除最近邻的样本μ值增加~40%
不同角度范围的μ值:
- 72° < θ < 90° (所有像素对):μ最大值~14%
- 72° < θ < 90° (排除最近邻):μ值更高,精度更好
- 60° < θ < 80°:μ最大值~8.5%,符合理论预期
关键发现:
- SBR与偏振调制因子μ呈明确正相关
- 在μ变化范围10.5%-14%内,SBR提升约2倍
- 该相关性表明μ可作为随机背景的估计器
高相关区vs低相关区:
- 70° < Δφ < 110° (理论高相关区):SBR最高
- -20° < Δφ < 20° (理论低相关区):SBR最低
- 验证了理论预期的偏振相关性空间分布
- Pryce-Ward理论:1947年建立的湮灭光子偏振相关理论基础
- 量子纠缠:近年来在MeV能区的量子纠缠研究进展
- 康普顿偏振测量:单层康普顿探测器技术发展
- 概念验证:多个研究组在实验室条件下的初步验证
- 探测器优化:GAGG:Ce晶体和SiPM技术的应用
- 系统集成:向临床应用的技术转化努力
本研究首次在临床相关活度范围内系统比较了偏振增强PET与传统PET的背景抑制性能,填补了从概念验证到实际应用之间的空白。
- 背景抑制优势:偏振相关康普顿事件在所有测试拓扑中均表现出比单像素事件更高的SBR
- 活度依赖性:较低活度下偏振增强效果更显著,为临床应用提供了重要参考
- μ因子的指示作用:首次建立了SBR与偏振调制因子的定量关系,μ可作为背景估计的新工具
- 技术可行性:在临床相关条件下验证了偏振增强PET的实用潜力
- 探测器几何:当前的平面几何限制了临床应用,需要发展环形几何
- 统计精度:较小的探测器阵列限制了统计精度和空间分辨率
- 复杂性增加:双像素事件处理增加了系统复杂性和计算负担
- 环境因素:正电子发射环境可能影响偏振相关性的测量
- 系统集成:开发完整的偏振增强PET扫描仪原型
- 算法优化:发展更高效的偏振相关事件重建算法
- 临床验证:在实际患者成像中验证技术优势
- 多模态融合:探索与其他成像技术的结合应用
- 创新性强:首次系统研究偏振增强PET的随机背景特性,开辟了新的研究方向
- 实验设计严谨:使用临床相关的活度范围,实验条件接近实际应用
- 理论与实验结合:很好地结合了量子力学理论和实际测量技术
- 数据分析全面:多角度分析了不同事件选择策略的影响
- 技术实用性:使用商业化组件构建系统,具有良好的可复现性
- 样本量限制:16×16像素阵列相对较小,限制了统计精度
- 几何局限:双模块平面几何与临床PET的环形几何存在差距
- 理论分析不足:对SBR与μ相关性的物理机制解释不够深入
- 对比基准:缺乏与商业PET系统的直接性能对比
- 成本效益分析:未讨论增加的系统复杂性与性能提升的权衡
学术价值:
- 为偏振增强PET领域提供了重要的实验基础数据
- 建立了新的背景评估方法和指标
- 推动了量子纠缠在医学成像中的应用研究
实用价值:
- 为下一代PET系统设计提供了技术路线
- 在高背景环境下的成像应用具有潜在优势
- 可能改善低活度示踪剂的成像质量
可复现性:
- 使用商业化组件,技术门槛相对较低
- 详细的实验参数和分析方法便于重复验证
- 开源分析软件(ROOT, Python)提高了可访问性
- 研究环境:核医学物理研究、医学成像技术开发
- 临床应用:低剂量PET成像、高背景环境成像
- 工业应用:核材料检测、安全检查等需要高信噪比的场景
- 教育培训:量子物理与医学成像结合的教学示范
本文引用了该领域的重要文献,包括:
- Pryce & Ward (1947): 湮灭光子角相关的理论基础
- Kuncic等 (2011): 偏振增强X射线成像的早期工作
- Watts等 (2021): MeV能区光子量子纠缠的实验验证
- Moskal等 (2025): 使用塑料PET扫描仪的非最大纠缠演示
总体评价:这是一篇在偏振增强PET领域具有重要意义的实验研究论文。虽然存在一些技术局限性,但其创新性的研究思路、严谨的实验设计和有价值的发现为该领域的发展做出了重要贡献。特别是SBR与偏振调制因子的相关性发现,为未来的系统优化提供了新的理论依据。