2025-11-25T22:43:18.712722

Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis

Siwal, Mohanty, Bose et al.
A systematic study is conducted to understand the coincident resolving time (CRT) for a pair of Lutetium-yttrium oxyorthosilicate (LYSO) and the plastic scintillation detector bars under the Geant4 framework. Crystals are coupled to a silicon photomultiplier single pad wafer with an appropriate optical coupling for signal readout. Pad reads the light photons undergoing optoelectronic conversion at the wafer site and generates electrical pulses with a bi-exponential shape. These signals are used to determine the trigger time stamp of back-to-back gamma-rays emitted from a point source, enabling the evaluation of CRT performance at different plastic scintillator lengths. For the LYSO detector, the simulation yields the CRT response of 174 ps, which aligns to the literature reported value for the dimensions 2 mm x 2 mm x 10 mm. To identify the plastic scintillator dimensions with an integrated gamma-ray detection efficiency comparable to LYSO's photopeak efficiency at 511 keV gamma photons, various bar lengths of commercial plastic BC404 and TIFR Ooty's in-house developed plastic material are attempted in Geant4. Consequently, for both the plastic scintillators, the equivalent length (for the same cross-sectional area) was found to be 4 times that of LYSO crystal length at a threshold of 25 keV. CRT value determined for this dimension is found to be $\approx$ 300 ps for both the plastic medium. It suggests that, if an animal preclinical PET scanner is developed with plastic bars, the minimum achievable image resolution (FWHM) of $\approx$ 4.5 cm can be expected for a pair of detection elements.
academic

Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis

基本信息

  • 论文ID: 2510.14437
  • 标题: Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis
  • 作者: Davinder Siwal, P. K. Mohanty, D. Bose, A. Jain
  • 所属机构:
    • Department of Physics, Central University of Kashmir
    • Tata Institute of Fundamental Research, Mumbai
  • 分类: physics.ins-det (仪器与探测器物理)
  • 发表时间: 2025年10月16日
  • 论文链接: https://arxiv.org/abs/2510.14437v1

摘要

本研究在Geant4框架下系统性地研究了LYSO(钇铥氧硅酸盐)和塑料闪烁体探测器条的符合时间分辨率(CRT)。晶体通过适当的光耦合与硅光电倍增管(SiPM)单片晶圆耦合进行信号读出。研究通过双指数形状的电脉冲确定点源发射的背靠背γ射线的触发时间戳,评估不同塑料闪烁体长度下的CRT性能。对于2×2×10 mm³的LYSO探测器,仿真得到174 ps的CRT响应,与文献报告值一致。为了确定与LYSO在511 keV γ光子光峰效率相当的塑料闪烁体尺寸,研究测试了商用BC404和TIFR Ooty内部开发塑料材料的各种条长度。结果表明,在25 keV阈值下,两种塑料闪烁体的等效长度(相同横截面积)是LYSO晶体长度的4倍,对应的CRT值约为300 ps。这表明如果使用塑料条开发动物临床前PET扫描仪,一对探测元件可实现的最小图像分辨率(FWHM)约为4.5 cm。

研究背景与动机

问题背景

  1. 传统PET扫描仪的挑战:传统动物PET扫描仪采用LYSO晶体,虽然具有优异的光输出、快速衰减时间和良好的能量分辨率,但成本昂贵,限制了其在发展中国家的应用。
  2. 塑料闪烁体的潜力:塑料闪烁体具有显著优势,包括:
    • 成本比基于镥的晶体低80倍
    • 体积可扩展性更好
    • 衰减长度约长100倍
    • 衰减时间比LYSO快一个数量级
  3. 技术可行性探索:Jagiellonian大学的最新研究重新激发了对塑料闪烁体在全身PET扫描仪中应用的兴趣,为本研究提供了动机。

研究意义

本研究旨在为印度研究环境下开发成本效益高的动物PET系统提供技术支撑,具有重要的实用价值和社会意义。

核心贡献

  1. 建立了完整的Geant4仿真框架:系统性地模拟了LYSO和塑料闪烁体探测器的光子传输、SiPM响应和信号处理过程。
  2. 验证了仿真模型的准确性:LYSO探测器的CRT仿真结果(174 ps)与文献报告值高度一致。
  3. 确定了塑料闪烁体的等效尺寸:发现塑料闪烁体需要4倍于LYSO的长度才能达到相当的探测效率。
  4. 评估了两种塑料闪烁体材料:对比分析了商用BC404和TIFR Ooty开发的塑料材料,发现性能相当。
  5. 提供了成本效益分析:为动物PET系统的实际应用提供了技术参数和性能预期。

方法详解

理论基础:光传输理论

研究基于Seifert等人建立的光子色散理论模型,采用多元高斯密度函数描述光子到达时间的概率分布:

P[ttannih]=1(2π)jΣexp(12(tμtannih)TΣ1(tμtannih))P[t | t_{annih}] = \frac{1}{\sqrt{(2\pi)^j |\Sigma|}} \cdot \exp\left(-\frac{1}{2}(t - \mu - t_{annih})^T\Sigma^{-1}(t - \mu - t_{annih})\right)

其中,Fisher信息矩阵为: I(tannih)=1TΣ11=i=1nj=1n(Σ1)ijI(t_{annih}) = 1^T\Sigma^{-1}1 = \sum_{i=1}^{n}\sum_{j=1}^{n}(\Sigma^{-1})_{ij}

Cramér-Rao下界为: σlb2=1I(tannih)\sigma_{lb}^2 = \frac{1}{I(t_{annih})}

Geant4仿真实现

几何建模

  • 探测器配置:一对闪烁体条放置在78 mm径向距离(典型动物PET扫描仪的机架半径)
  • 包装材料:50 μm厚的环氧树脂反射涂层包装
  • 光学耦合:200 μm光学润滑脂层 + 150 μm厚SiPM晶圆
  • 表面处理:考虑晶体表面粗糙度,采用微面角色散参数σα

材料属性设置

参数LYSOPlastic-OotyPlastic-BC404
密度 (g/cm³)7.11.0321.032
折射率1.811.581.58
吸收长度 (cm)1.2100160
光产额 (MeV⁻¹)270001000011832
衰减时间 (ns)431.51.5

多线程优化

采用G4MTRunManager实现24线程并行处理,显著减少了光子追踪的仿真时间。

SiPM信号建模

光子到达时间计算

tarriv=tγ+ΔtLt+tpst_{arriv} = t_\gamma + \Delta t_{Lt} + t_{ps}

其中:

  • tγt_\gamma:γ射线相互作用时间
  • ΔtLt\Delta t_{Lt}:光传输时间
  • tpst_{ps}:SiPM传播时间(高斯分布,σ = 80 ps)

单光电子响应函数

采用双指数脉冲模型: fSER=A[exp(ttarrivτd)exp(ttarrivτr)]f_{SER} = A \cdot \left[\exp\left(-\frac{t-t_{arriv}}{\tau_d}\right) - \exp\left(-\frac{t-t_{arriv}}{\tau_r}\right)\right]

参数设置:上升时间τr = 500 ps,衰减时间τd = 5 ns。

信号处理

  • 量子效率:35%
  • 能量分辨率:11%(511 keV)
  • 系统带宽:1 GHz
  • 电子学时间色散:57 ps (FWHM)

实验设置

仿真配置

  • Geant4版本:4.10.07.p02
  • 硬件平台:Intel 6核心(第12代Core i5)+ 16 GB RAM + Ubuntu 22.04.5 LTS
  • 并行处理:24线程
  • 事件数量:4,615个全能量符合γ射线事件
  • 光子统计:每事件至少300个光子到达SiPM

评价指标

  • 符合时间分辨率(CRT)CRT(FWHM)=2.35×σt12+σt22+σjitter2CRT(FWHM) = 2.35 \times \sqrt{\sigma_{t1}^2 + \sigma_{t2}^2 + \sigma_{jitter}^2}
  • 探测效率ϵCompton=康普顿连续谱事件数各向同性发射事件数\epsilon_{Compton} = \frac{\text{康普顿连续谱事件数}}{\text{各向同性发射事件数}}
  • 空间分辨率Δr=cΔt2\Delta r = \frac{c\Delta t}{2}

对比基准

  • LYSO晶体(2×2×10 mm³)作为参考标准
  • 文献报告的CRT值用于验证仿真准确性

实验结果

LYSO探测器验证

  • 仿真CRT结果:174 ps
  • 文献报告值:175 ps
  • 验证结论:仿真模型准确可靠

塑料闪烁体性能分析

光学时间色散

  • 塑料闪烁体的光学时间色散与条长度呈近线性关系
  • 由于更长的吸收长度,塑料材料的光子时间色散比LYSO更宽

CRT性能对比

条长度 (mm)BC404 CRT (ps)Ooty CRT (ps)
20~250~250
30~280~280
40~300~300
45~320~320

探测效率分析

  • 等效长度要求:塑料闪烁体需要40 mm长度才能达到与10 mm LYSO相当的探测效率(0.5×10⁻³%)
  • 效率-时间分辨率权衡:20 mm塑料条可达到与LYSO相当的时间分辨率,但探测效率仅为LYSO的40%

图像分辨率预测

使用40 mm塑料条的动物PET扫描仪预期最小图像分辨率(FWHM)约为4.5 cm。

相关工作

PET探测器技术发展

传统动物PET扫描仪主要采用LYSO晶体阵列,典型的CTR值为1-4 ns。相关研究报告:

  • 3×3×10 mm³ LYSO + SensL SiPM:190 ps CTR
  • 4 mm²横截面条:15 mm长度171.6 ps,20 mm长度179 ps

塑料闪烁体PET应用

Jagiellonian大学团队的开创性工作重新激发了塑料闪烁体在PET成像中的应用兴趣,特别是在全身PET扫描仪方面。

理论建模方法

本研究采用Seifert等人建立的光子色散理论,该方法为时间分辨率建立了下界,被广泛认可为标准理论框架。

结论与讨论

主要结论

  1. 技术可行性:塑料闪烁体可作为LYSO的替代方案用于动物PET扫描仪
  2. 性能权衡:40 mm塑料条可达到与LYSO相当的探测效率,但CRT降至300 ps
  3. 成本优势:塑料闪烁体的成本优势显著,适合资源受限环境
  4. 材料等效性:TIFR Ooty开发的塑料材料与商用BC404性能相当

局限性

  1. 空间分辨率限制:300 ps的CRT对应4.5 cm的图像分辨率,低于传统LYSO系统
  2. 仿真简化:未考虑SiPM的暗电流、光学串扰、余辉等效应
  3. 单一几何配置:仅研究了特定的探测器几何和配置

未来方向

  1. MOBY体模成像研究:使用2×2×40 mm³塑料条进行体模成像验证
  2. 机架设计优化:探索不同网格设计和机架配置以优化性能
  3. 实验验证:进行实际硬件测试验证仿真结果

深度评价

优点

  1. 方法学严谨:建立了完整的理论-仿真-验证框架,方法学可靠
  2. 实用价值高:为成本敏感的应用场景提供了可行的技术方案
  3. 仿真详尽:考虑了光传输、表面反射、SiPM响应等多个物理过程
  4. 验证充分:通过与文献值对比验证了仿真模型的准确性

不足

  1. 性能妥协明显:塑料闪烁体的时间分辨率和空间分辨率均低于LYSO
  2. 应用场景受限:4.5 cm的图像分辨率可能限制其在精细成像中的应用
  3. 缺乏实验验证:完全基于仿真,缺乏实际硬件测试数据
  4. 成本分析粗略:虽然提及成本优势,但缺乏详细的经济性分析

影响力

  1. 技术贡献:为PET探测器技术提供了新的设计思路
  2. 社会价值:有助于在资源受限地区推广PET成像技术
  3. 学术意义:为塑料闪烁体在医学成像中的应用提供了理论基础

适用场景

  1. 动物实验研究:适合对空间分辨率要求不高的动物代谢研究
  2. 教学培训:可用于PET成像原理的教学和培训
  3. 技术验证:作为新算法和重建技术的验证平台
  4. 资源受限环境:适合预算有限的研究机构和发展中国家

参考文献

论文引用了25篇重要参考文献,涵盖了PET成像、闪烁体物理、Geant4仿真等多个领域的关键工作,为研究提供了坚实的理论基础。


总体评价:这是一篇技术扎实、实用价值高的应用物理研究论文。虽然在性能上存在妥协,但为特定应用场景提供了有价值的技术方案,具有重要的社会意义和推广价值。