Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis
Siwal, Mohanty, Bose et al.
A systematic study is conducted to understand the coincident resolving time (CRT) for a pair of Lutetium-yttrium oxyorthosilicate (LYSO) and the plastic scintillation detector bars under the Geant4 framework. Crystals are coupled to a silicon photomultiplier single pad wafer with an appropriate optical coupling for signal readout. Pad reads the light photons undergoing optoelectronic conversion at the wafer site and generates electrical pulses with a bi-exponential shape. These signals are used to determine the trigger time stamp of back-to-back gamma-rays emitted from a point source, enabling the evaluation of CRT performance at different plastic scintillator lengths. For the LYSO detector, the simulation yields the CRT response of 174 ps, which aligns to the literature reported value for the dimensions 2 mm x 2 mm x 10 mm. To identify the plastic scintillator dimensions with an integrated gamma-ray detection efficiency comparable to LYSO's photopeak efficiency at 511 keV gamma photons, various bar lengths of commercial plastic BC404 and TIFR Ooty's in-house developed plastic material are attempted in Geant4. Consequently, for both the plastic scintillators, the equivalent length (for the same cross-sectional area) was found to be 4 times that of LYSO crystal length at a threshold of 25 keV. CRT value determined for this dimension is found to be $\approx$ 300 ps for both the plastic medium. It suggests that, if an animal preclinical PET scanner is developed with plastic bars, the minimum achievable image resolution (FWHM) of $\approx$ 4.5 cm can be expected for a pair of detection elements.
academic
Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis
This study systematically investigates the coincidence time resolution (CRT) of LYSO (lutetium yttrium oxyorthosilicate) and plastic scintillator detector bars within the Geant4 framework. Crystals are coupled to silicon photomultiplier (SiPM) monolithic wafers through appropriate optical coupling for signal readout. The research determines trigger timestamps of back-to-back γ-rays emitted from point sources through bi-exponential shaped electrical pulses, evaluating CRT performance across different plastic scintillator lengths. For a 2×2×10 mm³ LYSO detector, the simulation yields a CRT response of 174 ps, consistent with literature-reported values. To determine plastic scintillator dimensions equivalent to LYSO in 511 keV γ-photon photopeak efficiency, the study tests various bar lengths of commercial BC404 and TIFR Ooty internally developed plastic materials. Results demonstrate that at a 25 keV threshold, the equivalent length of both plastic scintillators (same cross-sectional area) is 4 times that of the LYSO crystal, with corresponding CRT values approximately 300 ps. This indicates that if plastic bars are used to develop animal preclinical PET scanners, the minimum image resolution (FWHM) achievable by a pair of detector elements would be approximately 4.5 cm.
Challenges of Traditional PET Scanners: Traditional animal PET scanners employ LYSO crystals, which despite exhibiting excellent light output, fast decay time, and good energy resolution, are prohibitively expensive, limiting their application in developing countries.
Potential of Plastic Scintillators: Plastic scintillators offer significant advantages, including:
Cost 80 times lower than lutetium-based crystals
Better volume scalability
Attenuation length approximately 100 times longer
Decay time one order of magnitude faster than LYSO
Exploration of Technical Feasibility: Recent research from Jagiellonian University has rekindled interest in applying plastic scintillators to whole-body PET scanners, providing motivation for this study.
This study aims to provide technical support for developing cost-effective animal PET systems within the Indian research environment, possessing important practical value and social significance.
Established a comprehensive Geant4 simulation framework: Systematically simulated photon transport, SiPM response, and signal processing for both LYSO and plastic scintillator detectors.
Validated simulation model accuracy: LYSO detector CRT simulation results (174 ps) show high consistency with literature-reported values.
Determined equivalent dimensions for plastic scintillators: Discovered that plastic scintillators require 4 times the length of LYSO to achieve comparable detection efficiency.
Evaluated two plastic scintillator materials: Comparative analysis of commercial BC404 and TIFR Ooty developed plastic materials, finding comparable performance.
Provided cost-benefit analysis: Supplied technical parameters and performance expectations for practical application of animal PET systems.
The research is based on the photon dispersion theory model established by Seifert et al., employing multivariate Gaussian density functions to describe the probability distribution of photon arrival times:
The pioneering work by the Jagiellonian University team has rekindled interest in plastic scintillator applications in PET imaging, particularly for whole-body PET scanners.
This study employs the photon dispersion theory established by Seifert et al., which establishes a lower bound for time resolution and is widely recognized as the standard theoretical framework.
The paper cites 25 important references covering multiple fields including PET imaging, scintillator physics, and Geant4 simulation, providing solid theoretical foundation for the research.
Overall Assessment: This is a technically sound and practically valuable applied physics research paper. While compromises exist in performance, it provides valuable technical solutions for specific application scenarios with important social significance and promotion potential.